Francis R. Patterson,Jennifer Thomson,Joseph Benevenia,Kathleen S. Beebe,Michael N. Kang, Biomechanical Study of Simulated Metastatic Peri-Acetabular Defects Journal of Surgical Oncology 2018 2674-3000 http://dx.doi.org/10.31487/j.JSO.2018.01.003 https://www.sciencerepository.org/biomechanical-study-of-simulated-metastatic-peri-acetabular-defects_JSO-1-103 Abstract: Peri-acetabular metastasis presents the risk of fracture. However, the relationship between bone loss and fracture risk is unknown. Utilizing fresh frozen intact cadaver pelves, Harrington class I lesions were simulated. Lesions were categorized as small (<40%) or large (>40%) Axial loading was applied in a quasistatic mechanism by a mechanical testing system to catastrophic failure, with the load-to-failure force being measured. There was statistical difference in the load-to-failure force between hemi-pelves with large defects and their matched controls. It was found that mean load-to-failure strength of a hemi-pelvis with a Class I defect was more than 2.3-2.45 times the calculated hip joint reactive force. Thus, it was concluded that a pelvis with a peri-acetabular defect of greater than 40% has a significantly lower load-to-failure strength compared to an intact pelvis. However, even this large Class I defect will fracture only at supraphysiologic loads and the hip joint may not normally be exposed to forces that high.Keywords: Peri-acetabular, Harrington, biomechanical