Table 2: Confusion matrix of
ventricular arrhythmia detection for 12 lead standard ECG and smartphone ECG in
comparison to cardiologist diagnosis for overall evaluation.
|
Parameters |
12 lead Gold standard |
Smartphone ECG |
|
True Positive |
30 |
46 |
|
True Negative |
1113 |
1169 |
|
False Positive |
32 |
38 |
|
False Negative |
162 |
84 |